如何在質(zhì)譜儀中重新啟動 EPC How do I reboot the EPC in my mass spectrometer?
How do I reboot the EPC in my mass spectrometer? -?
如何在質(zhì)譜儀中重新啟動 EPC? 菜鳥博士
ENVIRONMENT
環(huán)境
Xevo TQ-MS
Xevo TQD
Xevo?TQ-S?Xevo TQ-S
Xevo?TQ-XS?Xevo TQ-XS
Xevo TQ-S micro
ACQUITY?TQD?水資源總量
ACQUITY?SQD?ACQUITY SQD
ACQUITY QDa
3100 Mass Detector?3100質(zhì)量探測器
Quattro Micro
Quattro Premier
Xevo G2?Xevo g 2
Xevo G2-S
Xevo G2-XS
Synapt G2?Synapt g 2
Synapt G2-S
Synapt G2-Si
Synapt?XS?Synapt XS
ANSWER
答案
First use?Telnet to restart the EPC.?All?mass spectrometers have a reset button for the EPC, but if not done correctly, the system may vent.
首先使用 Telnet 重新啟動 EPC。所有的質(zhì)譜儀都有一個(gè) EPC 重置按鈕,但是如果操作不當(dāng),系統(tǒng)可能會泄漏。
Resetting the EPC?using Telnet:
使用 Telnet 重置 EPC:
Click Windows.?點(diǎn)擊窗口
In the search bar, type "Telnet epc"?and press Enter.?在搜索欄中,鍵入“ Telnet epc”,然后按回車
If Telnet is currently configured on your computer, the DOS prompt window launches.?如果 Telnet 當(dāng)前在您的計(jì)算機(jī)上配置,DOS 提示窗口就會啟動
The window should contain a ">", followed by a blinking cursor.?窗口應(yīng)該包含一個(gè)“ >”,后面跟著一個(gè)閃爍的光標(biāo)
Type "reboot" in lowercase and press Enter.?在小寫字母中鍵入“ reboot”,然后按 Enter 鍵
The EPC?is now rebooting, so you can close the Telnet window. Depending on the MS,?the reboot?can take up to five minutes.?A chime?should signal completion.?EPC 現(xiàn)在正在重新啟動,因此您可以關(guān)閉 Telnet 窗口。根據(jù) MS 的不同,重新啟動可能需要5分鐘。一聲鐘響應(yīng)表示完成