waters UPLC 如何校準針頭 z 軸 How to calibrate the needle Z axis
How to calibrate the needle Z axis - WKB28509
如何校準針頭 z 軸 -WKB28509
OBJECTIVE or GOAL
目的或目標
Calibrate the needle Z axis.
校準針 z 軸。
ENVIRONMENT
環(huán)境
ACQUITY UPLC Sample Manager (SM)?樣本管理器(SM)
ACQUITY UPLC H-Class?H 級
ACQUITY UPLC I-Class?一級水平
ACQUITY Arc Sample Manager FTN-R?弧樣本管理器
ACQUITY UPLC M-Class uSM-FL?M 級 uSM-FL
PROCEDURE
程序
Tip: Refer to the?diagnostics calibration video?to help you complete this task.
提示: 請參閱診斷校準視頻以幫助您完成此任務(wù)。
Remove the trays?and place a business card on the sample tray holder.?取出托盤,在樣品托盤架上放一張名片
From the console, select Maintain > Calibrate needle Z axis.?在控制臺上,選擇 Maintain > Calibrate needle z axis
In the Calibrate Needle Z Axis dialog box, click Start, and?then click OK.?在“校準針 z 軸”對話框中,單擊“開始”,然后單擊“確定”
Use the +Z button to drive the needle down to within 1 mm of the tray surface.?使用 + z 按鈕驅(qū)動針下降到1毫米以內(nèi)的托盤表面
Switch the displacement increment to 0.1 mm and lower the needle until it just barely touches the business card.?切換位移增量到0.1毫米和降低針,直到它剛剛接觸的名片
Click Save, and?then Yes?in the confirmation window.?單擊“保存”,然后在確認窗口中單擊“是”