【液相問題解決方案】儀器應(yīng)用waters PDA Lamp energy low(燈能量低)錯(cuò)誤

2022-11-26 12:30 作者:菜鳥博士_雜貨鋪  | 我要投稿

故障描述

  • 接通電源后初步診斷時(shí)出現(xiàn) Lamp energy low(燈能量低)錯(cuò)誤


環(huán)境

  • 2998 光電二極管陣列檢測器


原因
1填充檢測器池的溶劑具有 UV 吸光度。

2.超出燈使用壽命


解決方法
1.將低 UV 吸收的溶劑(如甲醇)輸送到流通池中,然后打開電源。

2. 更換燈。部件號(hào) 201000273


SYMPTOMS

  • “Lamp energy low” error occurs during initial diagnosis, after power-on

  • "Lamp energy low" message

  • Lamp has far exceeded the number of recommended hours

  • PDA Shutter not found error message.

  • Low lamp energy when check at 225 nm with water in flow cell.


ENVIRONMENT

  • 2998 Photodiode Array Detector


CAUSE
The solvent that fills the detector cell has UV absorption.

Lamp life exceeded

FIX or WORKAROUND
Deliver a solvent with low UV absorption, such as methanol, to the cell, and then power-on.

Replace the lamp. part number 201000273