273x28.58mm無縫管子 DN250/DN200/DN400/DN40/DN125
273x28.58mm無縫管子 DN250/DN200/DN400/DN40/DN125

5. Apparatus
5.1 Specimen Carrier, provided with two conical supports
for locating the specimen. The test length (that is, the distance
between the point of contact of the specimen with one support
and the point ofcontact ofthe specimen with the other support)
shall be known to within 60.005 in. (0.13 mm), and the line of
plane passing through the points of contact shall be horizontal.
The specimen carrier and supports shall hold the specimen
without constraint so that the curvature, due to its def l ection,
will follow a vertical plane passing through the line joining the
points of contact between specimen and supports. Suitable
means shall be used to ensure test length.
5.2 Micrometer—traveling microscope, or equivalent
device, so connected to the specimen carrier that expansion
during heating of the carrier or connecting parts will not cause
appreciable displacement of the measuring device with respect
to the supports.
5.3 Bath—A stirred liquid bath or uniformly heated enclo-
sure in which the specimen carrier, together with adjustable
electric heating source is placed. The specimen needs to be
maintained at the desired temperatures, with a variation in
temperature throughout the gage length of the specimen not to
exceed 0.5 % of the temperature range used in the test.
5.4 Temperature Measuring Apparatus, of such accuracy
that the individual temperatures shall be known to within
60.5°F (0.3°C).
5.5 Def l ection Index—Means shall be provided for measur-
ing the def l ection of the specimen at a point midway between
the points ofsupport and along the vertical line intersecting the
line joining the points of support. Such means may comprise a
transmission rod disposed with its axis vertical and terminating
in a point or knife-edge, which shall engage the specimen
midway between the points of contact with the supports.
5.5.1 The transmission rod shall be mounted in such a
manner that it is free to move in the direction of its axis. The
rod shall bear at its free end an index suitable for microscopical
observation, or else an electrical contact with which a microm-
eter will permit the changes of the def l ection of specimen to be
accurately observed. Alternatively, the def l ection of the mid-
point of the specimen may be directly observed by optical
means whose line of sight is horizontal and passes through the
vertical line through the midpoint of the specimen. A microm-
eter screw with extended spindle making direct contact with
the specimen may be used. In this case, electrical means shall
be provided that will indicate contact without signif i cant
disturbance of the specimen. The measurement of the position
of the midpoint of the test specimen shall be of such accuracy
that the individual positions at the test temperatures shall be
known to within 60.0002 in. (0.005 mm).
5.5.2 If a transmission rod is used, it and any attached parts
shall be ofsuch weight or so counterweighted that they will not
cause a def l ection greater than 1 % of the maximum to be
produced by the action of the thermostat metal alone. When
free, the thermostat metal assumed very nearly a circular
curvature. Concentrated loading at the center of the specimen
will cause the curvature to be other than circular and may cause
signif i cant errors in the evaluation of f l exivity. The location of
the line passing through the points of contact between speci-
men and supports shall be known, with reference to the scale of
the micrometer, to within 60.002 in. (0.05 mm).
5.6 All metallic components of the f l exivity apparatus
should be made of very low coefficient of thermal expansion
components. The recommended material is Invar. 3