【液相應(yīng)用方案】與儀器無法通訊No communication with instruments

2022-05-31 15:05 作者:菜鳥博士_雜貨鋪  | 我要投稿

菜鳥博士Caesar


  • Instruments show communication failure on the Run Samples window

  • Unable to restore communication by resetting from Run Samples/console

  • When cleared from DHCP, the instruments do not appear upon reboot

ENVIRONMENT

  • Alliance HPLC System

  • Acquity systems with standalone ethernet box installed


CAUSE Ethernet switch defective

FIX or WORKAROUND

  1. Replace the 8-port switch (Ethernet Switch Box, 8-port - 725000561).

  2. Turn off the LAC/E and all instruments connected, and then turn on the LAC/E and instruments.

  3. Confirm that communication is restored.


ADDITIONAL INFORMATION


與儀器無法通訊

故障描述

  • 儀器在 Run Samples(運行樣品)窗口中顯示通訊故障

  • 無法從 Run Samples(運行樣品)/控制臺重置來恢復(fù)通訊

  • 從 DHCP 清除后,儀器在重新啟動后不會顯示


環(huán)境

  • Alliance HPLC 系統(tǒng)

  • 安裝了獨立以太網(wǎng)盒的 ACQUITY 系統(tǒng)


原因 以太網(wǎng)交換機故障

解決方法

  1. 更換 8 端口交換機(8 端口以太網(wǎng)交換盒 - 725000561)。

  2. 關(guān)閉 LAC/E 和所有連接的儀器,然后再打開 LAC/E 和儀器。

  3. 確認通訊是否恢復(fù)。